4040+e_22.docx / 06.10.17
ROBERT RIELE GmbH & Co KG
Page 54
TIME: 08:44:12
25 GLUCOSE 13.03
21 HDL-C 367
27 UREA COL 197.2
29 CK-MB 1128.1
31 GOT 189.9
7.2.5.3 MONTHLY MEMORY
After query of method number select serum
1 or 2 of the method.
[S1] Select monthly memory for serum 1
[S2] Select monthly memory for serum 2
[EXIT] Return to previous window
In the overview window of the selected
method all data of quality control are visible.
In the line above the keys following
information the current reading is indicated:
(# 1)
Numerator of the monthly memory.
The oldest reading corresponds to the 1.
(01.27.06)
Date of reading
(13.30)
Reading
(-1.s)
Deviation of the reading lies within
minus 1s. From +/-3s the warning level
starts. With a deviation of > 3s an * is
displayed. For the calculation of the quality
values of a method at least 20 readings in
the month memory must be present!
(1)
User identification
The keys have following functions:
[+] Change to next reading
[-] Change to previous reading
[DEL] Delete all measuring data of the
monthly memory of the selected method and
confirm again (e.g. at change of serum)
[MORE] Change to output dialog
[EXIT] Return to QC menu
C O N T R O L S E R U M
E X I T S 1 S 2
M O N T H L Y M E M O R Y S 2
M E T H O D 2 5 G L U C O S E
S E R U M 1 L T - S Y S a b n o r m a l
L O T G 3 1 2
C O M P A N Y L A B O R + T E C H
D A T E M A Y 0 6
R E Q U I R E D 1 4 . 4 m m o l / l
R A N G E L : 1 2 . 1 O : 1 6 . 7
C A L C U L . s : 1 . 0 1 2 C V : 7 . 2 9 8
m : 1 3 . 8 6 0 n : 2 0
1 0 7 . 0 9 . 0 7 1 3 . 3 0 - 1 s 1
E X I T M O R E D E L E T E - +