43
Timer Reset
{CS Menu}
IF
changing timer values while the output is already ON.
ON
‘
’ timers will reset/begin after exiting from the menu.
OFF
‘
’ timers will begin when the output is next turned ON.
Notes:
Timer begins counting after enabling the RD60xx output.
(If the output was already enabled, the timer is already running. )
e.g.
If you:
Turn on the output, and wait more than 1 min.
Enter the menu and set the ‘Timer Reset’ to 1 min
Exit from menu (apply setting)
Then:
Output will immediately turn OFF.
(Because 1 min. has already passed since the output was
turned ON.)
Enabling this option forces the timer to reset when the settings are applied.
V-SET
(Set Voltage, in Volts)
Can Only be Enabled by Panel button
“
V-SET
”
Value of the Voltage Setpoint, in Volts, for when the unit is in
. It also represents an
upper limit for Voltage when in CC Mode.
Rotate
“
Encoder
”
Knob to desired value.
Use
“
◄
”
or
“
►
”
to switch selected multiplier (cursor location).
OR
Enter the desired value using the
Keypad
.
UART Baudrate
➥
Communication
Selectable Options:
9600 / 19200 / 38400 / 57600 / 115200
Only applies to the INTERNAL communication ‘
’.
(NOT the front panel USB)
IF
‘WIFI’ selected as ‘
’ setting:
Baud Rate is FIXED at 115200, regardless of value selected.