9. Calibrating the Balance
Since electronic balances are affected by gravity gravitational acceleration, they produce different
values in various locations. Therefore, before use, balances must be calibrated at the location where
they are installed. Calibration is also required after long periods without use, or if a balance begins to
produce inaccurate values.
Calibration of a balance, or “span adjustment,” is required to produce accurate measurements.
※
Span adjustment should be performed with the balance installed perfectly level and without any load
on the measurement pan.
1 Activating the span adjustment function
cal
Func
Key released
Function
Zero/Tare
Memory
On/Off
Target
Function
Zero/Tare
Memory
On/Off
Target
Pressed
continuously
Press and hold down the Function key.
Release the key when the display
changes from [
Func
] to [
Cal
].
2 Starting the span adjustment
on
0
cal
Release
simultaneously.
Press the Function key while
holding down the Zero/Tare key.
Function
Zero/Tare
Memory
On/Off
Target
Function
Zero/Tare
Memory
On/Off
Target
Press the Function key while holding
down the Zero/Tare key, then release
both keys simultaneously. The display
flashes [
on 0
], indicating that
zero-point calibration is underway.
3 Zero
calibration
on
f.s
on
0
Function
Zero/Tare
Memory
On/Off
Target
Zero-point calibration is finished when
the display changes to [
on
F.S
].
Proceed to the calibration of the capacity
point.