8.2 Setup of Limit Values by Actual Quantity Loads
1 Starting the limit function
L. SET
Key released
Function
Zero/Tare
Memory
On/Off
Target
Continuous Pressing
Function
Zero/Tare
Memory
On/Off
Target
Press and hold down the Target key.
Release the key when [
L
.
SEt
] is
displayed.
The currently set lower-limit value
flashes.
2 Loading the sample for the lower-limit value
◄
0.00
Load the sample of the lower-limit value
on the measurement pan.
3 Saving the lower-limit value
◄
93.83
Function
Zero/Tare
Memory
On/Off
Target
Press the Memory key.
After the lower-limit value has been
saved, the balance displays it briefly and
proceeds to the following setup.
※
If One-point setup was chosen, the
setup is complete.
The display now changes to
[
H. SET
], indicating that the upper-limit
value can be set.
The currently set upper-limit value
flashes.
4 The
upper-limit
value setup
H. SET
◄
0.00
5 Loading the sample of the upper-limit value
◄
0.00
Load the sample of the upper-limit value
on the measurement pan.
6 Saving the upper-limit value
◄
158.48
Function
Zero/Tare
Memory
On/Off
Target
Press the Memory key.
After saving the upper-limit value, the
balance displays it briefly and terminates
the setup.
25