4.2 Description of Functions
Item Set
Value
Description
Bar graph display
1
.
.b
.
G
.
.
0
Disable
☆
1
Enable
Limit function
2
.
SEL
☆
0
Disable
1
Enable
Displayed only when limit
function is activated
Judgement
condition
21
.
Co
.
☆
1
Always judge (judges even when the balance is unstable)
2
Judge only when the balance is stable (does not judge
if the balance is unstable)
Judgement
range
22
.
L
0
Ranges 5 graduation is judged (5
graduation or below, including negative ranges, are not
judged.)
☆
1
The entire range is judged (the entire range, including
the negative, is judged).
Number of
points for
judgement
23
.
P 1
One-point setup (judges between OK and LO)
☆
2
Upper-limit and lower-limit values are set up (judges
among HI, OK and LO).
Auto-zero
(zero-tracking)
3
.
A
.
O
0
Disable
This function automatically sets the zero point
exactly to zero to prevent slight deviations.
☆
1
Enable
Auto power-off
4
.
A
.
P
.
0
Disable (balance operates
continuously)
This function is available
only when the balance is
battery-operated.
☆
1
Enable (balance powers off in
approximately three minutes)
Response speed
5
.
rE
.
0
Measurement by consecutive weighings.
1
Fast
↓
Slow
2
☆
3
4
5
Stability parameters
6
.
5
.
d
.
1
Wide (mild)
↓
Narrow (strict)
☆
2
3
4
Interface
7
.
1
.
F
.
0
Disable input/output
Six-digit numeric format
☆
1
2
Seven-digit numeric format
Setup of units of
measurement to be
displayed
Register selected
measuring units with
Function key.
☆
1 01
[
g
]
[ ] (ct)
☆
2 14
[ ] (oz)
15
[ ] (lb)
16
[ ] (ozt)
17
[ ] (dwt)
18
*
19
[
►
Lower right
] (grain)
Items marked
☆
are the default factory settings.
☆
1
~
☆
5: default settings [
81
.
5
.
u
]
~
[
85
.
5
.
u
.]
*
Not available in TC-12K.
13
Summary of Contents for TC-1200