Configuration
© Rice Lake Weighing Systems
●
All Rights Reserved
31
4.4.5 Setup – Program Menu
Figure 4-12. Setup – Program Menu
Parameter
Description
pwrupm
Power Up Mode – When the indicator is turned on, it performs a display test, then enters a warm up period;
Settings:
GO
(default)
– Performs display test, then enters weigh mode after brief warm up period
DELAY
– Performs display test, then enters a warm up period of 30 seconds
• If no motion is detected during warm up period, indicator enters weigh mode when warm up period ends
• If motion is detected, the 30 second timer is reset and the warm up period is repeated
regula
Regulatory Mode – Specifies the regulatory agency having jurisdiction over the scale site; the value specified for this parameter
affects the function of the front panel tare and zero keys; Settings:
NTEP
(default), OIML, CANADA, INDUST, NONE
• OIML, NTEP, and CANADA modes allow a tare to be acquired at a weight greater than zero; NONE allows tares to be
acquired at any weight value
• OIML, NTEP, and CANADA modes allow a tare to be cleared only if the gross weight is at no load; NONE allows tares
to be cleared at any weight value
• NTEP and OIML modes allow a new tare to be acquired even if a tare is already present; In CANADA mode, the
previous tare must be cleared before a new tare can be acquired
• NONE, NTEP and CANADA modes allow the scale to be zeroed in either gross or net mode as long as the current
weight is within the specified ZRANGE; In OIML mode, the scale must be in gross mode before it can be zeroed;
pressing the ZERO key in net mode clears the tare
• INDUST provides a set of sub-parameters to allow customization of tare, clear, and print functions in non Legal-for-
Trade scale installations
indset
Industrial Settings – Displays when REGULA parameter is set to INDUST; See
consnu
Consecutive Numbering – Allows sequential numbering for print operations; the value is incremented following each print
operation which includes <CN> in the ticket format;
Enter value:
0–9999999,
0
(default)
constu
Consecutive Number Startup Value – Specifies the initial consecutive number (CONSNU) value used when the consecutive
number is reset by sending CLRCN digital input;
Enter value:
0–9999999,
0
(default)
uid
Unit ID – Specifies the unit identification number with an alphanumeric value;
Enter characters: Up to 6 alphanumeric
characters,
1
(default)
accum
Accumulator – Accumulation can be toggled ON/OFF; if ON, accumulation occurs on print operation; if OFF, an accumulation
does not occur;
Settings:
OFF
(default), ON
Wmtthrh
Weighment Threshold – When the weight falls below the value set, the accumulator is rearmed;
Enter value: 0.0–9999999.0,
1000.0
(default)
Numwegh
Number of Weighments – Displays the total number of weighments;
Read only
Maxwegh
Maximum Weighment – Displays the maximum weighment allowed;
Read only
maxdate
Date/Time of Max Weight – Displays the date the maximum weighment occurred;
Read only
date
Date – Allows setting of the date format and date separator character
DATEFMT – Date Format;
Settings:
MMDDYY
(default), DDMMYY, YYMMDD, YYDDMM
DATESEP – Date Separator;
Settings:
SLASH
(default), DASH, SEMI, DOT
Table 4-11. Setup – Program Menu Parameters
Setup
Progrm
Pwrupm
Regula
indset
Datefmt
Datesep
Date
Timefmt
Timesep
Time
Locale
latorig
eleorig
latdest
eledest
Displays if
Locale
is set to
on
Displays if
regulat
is set to
indust
Consnu
Uid
Accum
constu
wmtthrh
maxwegh
maxdate
numwegh