Section 10
Integra 3™ Troubleshooting
58
Research Instruments Ltd
10
Alarm
Condition
Code
Message
Priority
Fault
Description/
Alarm Criteria
Alarm
Actions
Solution
E15
HEATED INSERT
TEMPERATURE
OUTSIDE ±0.5°C
Medium
Heated Stage
Insert temperature
has deviated by
more than 0.5°C
from the setpoint
temperature.
Alarm enabled
5 minutes after
the temperature
reaches ±2.5°
from the setpoint
temperature.
The Heated
Stage Insert
temperature
controller
continues to
operate.
This may be
caused by
placing either
hot or cold
objects on the
stage. In this
case either
remove the
object or wait
a short time for
the setpoint
temperature to
reached.
Sudden air
movements or
temperature
change can also
cause minor
temperature
fluctuations.
In this case
wait a short
while for the
temperature
controller to
respond.
Switch off the
Integra 3 and
verify that the
Heated Stage
Insert cables
and connector
are secure and
undamaged.
Possible heater
or controller
fault.