Section 8
Integra 3™ Alarms and System Status
38
Research Instruments Ltd
8
SECTION 8 - ALARMS AND SYSTEM STATUS
Off -
Please wait (Initial power up/setpoint/mode/calibration
changed) light will be off until temperature of all heating systems is
within ±0.5°C of setpoint.
Green (constantly on) -
Ready for use. Temperature of all heating
systems within ±0.5°C of setpoint.
Yellow (constantly on) -
Low Priority Alarm. Press the
Event Log
icon shown on the main screen to view specific alarm conditions.
Yellow (flashing) -
Medium Priority Alarm. Press the
Event Log
icon
shown on the main screen to view specific alarm conditions.
The status of the temperature control system is shown by the Status
Indicator Light on the left hand side of the Integra 3. Individual
temperatures are shown on the touch screen display.
Status Indicator Light
Note:
For a full list of possible faults relating to each alarm condition and
applicability of each alarm condition, refer to the Troubleshooting section
of this manual.
Once activated, all alarms are sounded immediately as the alarm
condition occurs.
Audible Alarms
Audible alarms are sounded to indicate Low and Medium Priority Alarms,
as described above. When an alarm is sounding, the alarm can be muted
by pressing the
Audio Alarm Off
icon which appears on the display (alarm
audio is automatically un-muted each time a new alarm becomes active).
When multiple alarms are active, the audio for the highest priority alarm
will be sounded. The alarm volume is not adjustable. Low and medium
priority alarms are sounded at the same level (this is a higher level than
other available indications used by the Integra 3).
In addition to the alarms listed above, sounds are emitted to indicate
touch screen key presses and that the toolholders are reaching the limits
of their vertical adjustment.
When multiple alarms are active, the icon and Status Indicator Light for
the highest priority alarm will be shown.