Section 10
Integra 3™ Troubleshooting
54
Research Instruments Ltd
10
Alarm
Condition
Code
Message
Priority
Fault
Description/
Alarm Criteria
Alarm
Actions
Solution
E06
HEATED INSERT
SENSOR FAILURE
Low /
Medium*
No signal/out of
range signal from
Heated Stage Insert
temperature sensor.
Alarm activates
at any time if the
temperature sensor
circuit fails to read a
valid temperature.
Temperature
is shown as
-.-°C. Heated
Stage Insert
controller
power is
set to 0%
until a valid
temperature
is read.
Switch off the
Integra 3 and
verify that the
Heated Stage
Insert cables
and connector
are secure and
undamaged.
Possible heater
or controller
fault.
E07
HEATED STAGE
OVER
TEMPERATURE
Medium
Heated Stage
Plate temperature
has exceeded
the maximum
allowable setpoint
temperature.
Alarm activates
at any time if the
temperature sensor
exceeds 46°C.
Heated
Stage Plate
controller
power is set
to 0% until
temperature
falls to
below 46°C.
Temperature
is shown as
-.-°C above
55°C.
Switch off the
Integra 3 and
verify that the
Heated Stage
Plate cables
and connectors
are secure and
undamaged.
Possible heater
or controller
fault.
E08
HEATED STAGE
TEMPERATURE
OUTSIDE ±2.5°C
Medium
Heated Stage Plate
temperature has
deviated more
than 2.5°C away
from the setpoint
temperature. Alarm
enabled 1 minutes
after reaching
±2.5°C from the
setpoint.
The Heated
Stage Plate
temperature
controller
continues to
operate.
Switch off the
Integra 3 and
verify that the
Heated Stage
Plate cables
and connectors
are secure and
undamaged.
Possible heater
or controller
fault.
* Low priority is set if condition appears within the first 5 seconds from power up. Medium priority is set if
condition appears after the first 5 seconds from power up.