- 4 -
EN
EN
4.6 Operating and display
functions
see also Figure 3
Action
Setting
Result
Display
switch on unit
• press on/off key
• unit is ready for
operation
• on/off LED is on
switch off the unit
• press on/off key
• unit is switched off
• all displays are off
start ultrasound
- now -
• select period by turn-
ing knob for cleaning
period
• press key ►■
(ultra-sound)
• ultrasound is
operating
• ultrasound LED is on
• set period LED is on
• remaining period LED blinks
(only in timer operation)
start ultrasound
- temperature-controlled*);
with mixing of cleaning liquid
*) if set temperature > actual
temperature
• set period
• set temperature
by turning knob for
temperature
• keep key ►■
pressed for > 2 sec
• heating operates
• ultrasound is started
automatically after
reaching the set tem-
perature
• set period ultrasound
runs down
• ultrasound LED blinks
• set period LED blinks
• as soon as the set temperature is
reached the ultrasound LED is on
• set period LED is on
• remaining period LED blinks
stop ultrasound before end
of set period
• turn set period to 0
or
• press key ►■
• ultrasound switched
off
• ultrasound LED is off
• set period LED is on
switch on heating
• select set
temperature
• heating operates
• set temperature LED is on
• actual temperature LED blinks and
goes towards set temp.
• as soon as actual temp. = set temp.,
only the set temp. LED is on
• if actual temperature > set tem
-
perature, the temperature LED starts
blinking again
switch off heating by hand
• turn set temperature
to position „0“
• heating switched off
• set temperature LED blinks
switch on sweep function*)
*) sweep and degas cannot be
operated at the same time
• select set period
• press key ►■
• press key sweep
• ultrasound operates
in sweep mode
• sweep LED is on
• ultrasound LED is on
• set period LED is on
• remaining period LED blinks
switch off sweep function
• press key sweep
• sweep function is
switched off
• ultrasound continues
in standard operating
mode
• sweep LED is off
• ultrasound LED is on
• set period LED is on
• remaining period LED blinks
switch on degas function*)
*) sweep and degas cannot be
operated at the same time
• select set period
• press key ►■
• press key degas
• ultrasound operates
in degas mode
• degas LED is on
• ultrasound LED is on
• set period LED is on
• remaining period LED blinks
switch off degas function
• press key degas
• degas function is
switched off
• ultrasound continues
in standard operating
mode
• degas LED is off
• ultrasound LED is on
• set period LED is on
• remaining period LED blinks
switch on auto degas func-
tion*)
*) sweep and degas cannot be
operated at the same time
• press key ►■
• keep degas key
pressed > 2 sec
• ultrasound operates
in auto degas mode
for 10 minutes and
then switches off
• degas LED blinks
• ultrasound LED blinks