2.4.9 Frequency Modulation
2.4.9.1 FM Deviation Setting
Test setup
Test setup 1 with modulation analyzer
Settings on SM300
LEVEL 0 dBm
FREQ 1 GHz
MODULATION FM DEVIATION 100 kHz
LFGenFreq 1 kHz
Setting on
modulation analyzer
Demodulation: FM
Detector: peak detector
Filter: 20 Hz to 23 kHz
Measurement
Read frequency deviation on modulation analyzer
2.4.9.2 FM Frequency Response
Test setup
Test setup 1 with modulation analyzer
Settings on SM300
LEVEL 0 dBm
FREQ 1 GHz
Deviation 50 kHz
Setting on
modulation analyzer
Demodulation: FM
Detector: peak detector
Measurement
The modulation frequency response is determined by varying the generator
frequency of the internal LF generator in the FM menu from 20 Hz to 20 kHz.
It is obtained from the difference between the lowest and highest measured
deviation.
2.4.9.3 FM Distortion
Test setup
Test setup 1 with modulation analyzer
Settings on SM300
LEVEL 0 dBm
FREQ 777 MHz
Deviation 50 kHz
LFGenFreq 1 kHz
Setting on
modulation analyzer
Demodulation: FM
Detector: peak detector
Audio: distortion
Measurement
Read distortion on modulation analyzer.
2.4.9.4 Output Reflection Coefficient
Test setup
Test setup 1 (use SWR bridge instead of the block mixer, preamplifier; the
scope is not used).