Remote control commands
R&S
®
NRPxxA(N)
90
User Manual 1177.6017.02 ─ 08
Manual operation:
See
9.8.3
Configuring corrections
It is possible to set some parameters that compensate for a change of the measured
signal due to fixed external influences.
.............................................................................................90
........................................................................................... 91
..............................................................................................92
Using the S-Parameters program
........................................................................... 95
9.8.3.1
Duty cycle corrections
The duty cycle is the percentage of one period during which the signal is active, when
pulse-modulated signals are corrected. The duty cycle is only evaluated in the continu-
ous average mode.
Remote commands:
[SENSe<Sensor>:]CORRection:DCYCle
............................................................................ 90
[SENSe<Sensor>:]CORRection:DCYCle:STATe
.................................................................. 90
[SENSe<Sensor>:]CORRection:DCYCle
<duty_cycle>
Available in continuous average mode.
Sets the duty cycle for measuring pulse-modulated signals. The duty cycle defines the
percentage of one period during which the signal is active. If the duty cycle is enabled,
the R&S NRPxxA(N) considers this percentage when calculating the signal pulse
power from the average power.
Parameters:
<duty_cycle>
Range:
0.001 to 100.00
*RST:
1.00
Default unit: Percent
Manual operation:
See
[SENSe<Sensor>:]CORRection:DCYCle:STATe
<state>
Enables or disables the duty cycle correction for the measured value.
Parameters:
<state>
*RST:
OFF
Manual operation:
See
Configuring basic measurement parameters