Remote control commands
R&S
®
NRPxxA(N)
92
User Manual 1177.6017.02 ─ 08
of S-parameter data for the 2
-
port device in the frequency range required by the appli-
cation.
Configuring the S
-
parameter correction
[SENSe<Sensor>:]CORRection:SPDevice:LIST?
................................................................ 92
[SENSe<Sensor>:]CORRection:SPDevice:SELect
.............................................................. 92
[SENSe<Sensor>:]CORRection:SPDevice:STATe
............................................................... 92
[SENSe<Sensor>:]CORRection:SPDevice:LIST?
Queries the list of the S-parameter data sets that have been downloaded to the power
sensor. The result of the query indicates the consecutive number and mnemonic of
each data set.
Usage:
Query only
[SENSe<Sensor>:]CORRection:SPDevice:SELect
<num>
Selects a downloaded data set for S-parameter correction.
See also
Parameters:
<num>
Range:
1 to 1999
*RST:
1; can differ if a calibration set defines another
value.
[SENSe<Sensor>:]CORRection:SPDevice:STATe
<state>
Enables or disables the S-parameter correction. If activated, uses the S-parameter
data set selected by
[SENSe<Sensor>:]CORRection:SPDevice:SELect
See also
Parameters:
<state>
*RST:
OFF; can differ if a calibration set defines another
value.
Example:
CORR:SPD:SEL 1
Selects an S-parameter correction data set.
CORR:SPD:STAT ON
Enables the S-parameter correction.
9.8.3.4
S-gamma corrections
Using the complex reflection coefficient, you can determine the power P delivered by
the signal source with considerably greater accuracy.
The coefficient of the signal source
Γ
source
is defined by its magnitude and phase:
●
[SENSe<Sensor>:]SGAMma:MAGNitude
●
Configuring basic measurement parameters