Remote control commands
R&S
®
NRPxxA(N)
94
User Manual 1177.6017.02 ─ 08
[SENSe<Sensor>:]SGAMma:MAGNitude
<magnitude>
Sets the magnitude of the complex reflection coefficient of the source,
Γ
source
.
Parameters:
<magnitude>
0.0
≙
ideal matched source
1.0
≙
total reflection
Range:
0.0 to 1.0
*RST:
0.0
Manual operation:
See
[SENSe<Sensor>:]SGAMma:PHASe
<phase>
Sets the phase angle of the complex reflection coefficient of the source,
Γ
source
.
Parameters:
<phase>
Range:
-360.0 to 360.0
*RST:
0.0
Manual operation:
See
9.8.3.5
I-gamma queries
For the current frequency, queries the complex input reflection coefficient
Γ
in
of the fol-
lowing:
●
Power sensor if the S-parameter correction is disabled.
(
[SENSe<Sensor>:]SGAMma:CORRection:STATe
OFF
)
●
S-parameter device if the S-parameter correction is enabled, S11 in
(
[SENSe<Sensor>:]SGAMma:CORRection:STATe
ON
)
[SENSe<Sensor>:]IGAMma:MAGNitude?
...........................................................................94
.................................................................................94
[SENSe<Sensor>:]IGAMma:EUNCertainty?
........................................................................95
[SENSe<Sensor>:]IGAMma:MAGNitude?
Queries the magnitude of the complex input reflection coefficient
Γ
in
.
Example:
IGAM:MAGN?
Query
1.739179E-02
Response
Usage:
Query only
[SENSe<Sensor>:]IGAMma:PHASe?
Queries the phase angle of the complex input reflection coefficient
Γ
in
. The result is
provided in degrees. Range:
-
180
degrees to +180
degrees.
Configuring basic measurement parameters