Browser-based user interface
R&S
®
NRPxxA(N)
42
User Manual 1177.6017.02 ─ 08
Watts, dBm or dBµV. To change the unit, you must specify the desired value together
with the full new unit once.
Example:
To change the representation of a "Trigger Level" of 100
µ
W into dBm, enter
-10dbm
in
the "Trigger Level" field. All future entries of solely numbers represent the value in
dBm. If you enter
-15
in the field, the "Trigger Level" value is set to -15.00
dBm.
If you want to revert the value to Watts, enter
50uW
. The "Trigger Level" value is set a
value of 50.00
µ
W, thus changing the unit for the further numeric entries.
6.3
Common settings
Describes the common sensor settings that are available for all measurement modes.
Access: main dialog of the web user interface > top pane
Measurement
Enables or disables the measurement.
Remote command:
Frequency
Sets the carrier frequency of the applied signal. This value is used for frequency-
response correction of the measurement result.
Remote command:
Offset
Groups the offset settings.
<State> ← Offset
Enables or disables the usage of the level offset.
Remote command:
[SENSe<Sensor>:]CORRection:OFFSet:STATe
Common settings