Remote control commands
R&S
®
NRPxxA(N)
66
User Manual 1177.6017.02 ─ 08
9.4.4
Configuring the measured values
Before starting a measurement, you can configure the measurand or enable the mea-
surement of additional-measured values.
..............................................................................................66
CALCulate:FEED
<mode>
If you query measurement data using
FETCh<Sensor>[:SCALar][:POWer][:
, the power sensor returns data of the measurand that was configured before.
Generally, this measurand is the average power. However, the power sensor can also
output data of other measurands.
To configure which measurand the
FETCh<Sensor>[:SCALar][:POWer][:AVG]?
command reads, use the
CALCulate:FEED
command before the measurement is ini-
tiated. Depending on the measurement mode, the following settings are possible:
SENS:FUNC
Possible CALC:FEED
Meaning
"POWer:AVG"
"POWer:AVERage"
Average value
"POWer:PEAK"
Peak value
"POWer:RANDom"
Randomly selected value from the
measurement interval
Parameters:
<mode>
*RST:
"POWer:AVERage"
[SENSe<Sensor>:]AUXiliary
<mode>
Enables the measurement of additional-measured values that are determined together
with the main-measured value.
Parameters:
<mode>
NONE | MINMax | RNDMax
NONE
No additional values are measured.
MINMax
Minima and maxima of the trace are transmitted together with
the measured value.
Usually, extreme values are lost due to averaging the measured
values.
RNDMax
Randomly selected samples are transmitted. All evaluations use
these values instead of the average values.
*RST:
NONE
Preparing for the measurement