![R&S NRP Series User Manual Download Page 115](http://html1.mh-extra.com/html/rands/nrp-series/nrp-series_user-manual_774882115.webp)
Remote control commands
R&S
®
NRPxxA(N)
115
User Manual 1177.6017.02 ─ 08
SYSTem:PRESet
Resets the power sensor.
The command essentially corresponds to the
*RST
command, with the exception that
the settings of the following commands are persistently held:
INITiate:CONTinuous
SENSe:AVERage:TCONtrol
Usage:
Event
SYSTem:REBoot
Reboots the power sensor.
Usage:
Event
SYSTem:RESTart
Restarts the firmware of the power sensor.
Usage:
Event
SYSTem:RUTime
<update_time>
Effective only in the NRP legacy mode. Relevant only in continuous measurement
mode,
Sets the result update time. That is the maximum rate in which the power sensor can
output measurement results.
Parameters:
<update_time>
Range:
0.0 to 10.0
*RST:
0.1
Default unit: Seconds
SYSTem:SERRor?
Queries the next static error, if available.
Static errors, as a rule, prevent the execution of normal measurements.
Errors in SCPI communication are queried using
.
Usage:
Query only
SYSTem:SERRor:LIST:ALL?
Queries all changes in the static error queue that have not been read yet and removes
them from the queue.
Configuring the system