![background image](http://html1.mh-extra.com/html/rands/1419-0006k02/1419-0006k02_user-manual_774879138.webp)
Remote control commands
R&S
®
NRPxxS(N)
138
User Manual 1177.5079.02 ─ 15
The power sensor performs the change of the hostname immediately after the com-
mand is sent. For this purpose, the power sensor restarts its connection to the network,
which can take several seconds. During this time, you cannot address the power sen-
sor. After the restart, you can only address the power sensor using the newly set host-
name.
Note:
It is recommended that you do not change the default hostname to avoid prob-
lems with the network connection. However, if you change the hostname, be sure to
use a unique name.
Parameters:
<hostname>
Example:
SYST:COMM:NETW:COMM:HOST
'powersensor-2nd-floor'
Sets
powersensor-2nd-floor
as new hostname.
SYSTem:COMMunicate:NETWork:IPADdress
<ipaddress>
Effective only:
●
For the R&S
NRP LAN power sensors.
●
SYSTem:COMMunicate:NETWork:IPADdress:MODE
is set to
STATic
.
Sets the IP address of the sensor.
Parameters:
<ipaddress>
Example:
SYST:COMM:NETW:IPAD '192.168.10.29'
Sets
192.168.10.29
as IP address.
Manual operation:
See
SYSTem:COMMunicate:NETWork:IPADdress:GATeway
<gateway>
Effective only:
●
For the R&S
NRP LAN power sensors.
●
SYSTem:COMMunicate:NETWork:IPADdress:MODE
is set to
STATic
.
Sets the IP address of the default gateway.
Parameters:
<gateway>
Example:
SYST:COMM:NETW:IPAD:GAT '192.168.10.254'
Sets
192.168.10.254
as IP address of the default gateway.
Manual operation:
See
SYSTem:COMMunicate:NETWork:IPADdress:INFO?
Effective only for the R&S
NRP LAN power sensors.
Configuring the system