Remote control commands
R&S
®
NRPxxS(N)
101
User Manual 1177.5079.02 ─ 15
measurands have been activated before exe-
cuting the measurement, the user data content is finished here. In case that auxiliary
measurands have been selected, the above section is repeated for every auxiliary
measurand. The user data content looks like:
AVGf3100...(400byte AVG values)...MINf3100...(400byte min.
values)...MAXf3100...(400byte max. values)...
Where each of
...(400byte AVG values)...
...(400byte min. values)...
...(400byte max. values)...
Stands for 400
bytes as the equivalent of 100
float values.
The user data content is embedded in the arbitrary block response data response.
Example:
TRAC:DATA?
Usage:
Query only
[SENSe<Sensor>:]TRACe:MPWidth?
Queries the attainable time resolution in trace mode. The result is the smallest possible
distance between two pixels, i.e. it is the smallest time interval that can be assigned to
a pixel.
Usage:
Query only
[SENSe<Sensor>:]TRACe:OFFSet:TIME
<time>
Adds an offset to the beginning of the trace sequence. Thus, the trace in the result dis-
play is moved in positive or negative x-direction. If you measure with more than one
sensor, you can use this offset to arrange the traces to each other. The start of record-
ing relative to the trigger event is set using
.
Parameters:
<time>
Range:
- (500 ms + TRGger:DELay) to 100 s
*RST:
0.0
Default unit: Seconds
Example:
TRAC:OFFS:TIME 1.0
Manual operation:
See
[SENSe<Sensor>:]TRACe:POINts
<points>
Sets the number of required values per trace sequence.
Parameters:
<points>
Range:
1 to 100000
*RST:
260
Configuring the measurement modes