Remote control commands
R&S
®
NRPxxS(N)
104
User Manual 1177.5079.02 ─ 15
Parameters:
<maximum_time>
Range:
0.01 to 999.99
*RST:
4.00
Default unit: Seconds
Manual operation:
See
[SENSe<Sensor>:]AVERage:COUNt:AUTO:NSRatio
<nsr>
Determines the relative noise component in the measurement result for the measure-
ment modes with scalar results. These measurement modes arecontinuous average,
burst average and timeslot, provided the particular sensor supports them.
This command is only effective if the auto average calculation is enabled:
●
[SENSe<Sensor>:]AVERage:COUNt:AUTO
ON
●
[SENSe<Sensor>:]AVERage:COUNt:AUTO:TYPE
NSR
The noise component is defined as the magnitude of the level variation in dB caused
by the inherent noise of the sensor (two standard deviations).
The query returns the relative noise component in the measured value.
Parameters:
<nsr>
Range:
100.000e-6 to 1.00
*RST:
0.01
Default unit: dB
Manual operation:
See
[SENSe<Sensor>:]AVERage:COUNt:AUTO:RESolution
<resolution>
Defines the number of significant places for linear units and the number of decimal pla-
ces for logarithmic units that are likely free of noise in the measurement result.
The setting is only considered, if the following applies:
●
[SENSe<Sensor>:]AVERage:COUNt:AUTO
ON
●
[SENSe<Sensor>:]AVERage:COUNt:AUTO:TYPE
RES
Parameters:
<resolution>
Range:
1 to 4
*RST:
3
Manual operation:
See
[SENSe<Sensor>:]AVERage:COUNt:AUTO:SLOT
<slot>
Available only in timeslot mode.
Sets a timeslot from which the measured value is used to determine the filter length
automatically. The timeslot number must not exceed the number of the currently set
timeslots.
Configuring basic measurement parameters