CALibration<Channel>:ZERO:AUTO
................................................................................126
CALibration:DATA
<caldata>
Writes a binary calibration data set in the memory of the power sensor.
Parameters:
<caldata>
<block_data>
CALibration:DATA:LENGth?
Queries the length in bytes of the calibration data set currently stored in the flash mem-
ory. Programs that read out the calibration data set can use this information to deter-
mine the capacity of the buffer memory required.
Example:
CAL:DATA:LENG?
Query
57392
Response
Usage:
Query only
CALibration:USER:DATA
<caldata>
Transfers the user calibration data set, which mainly contains S-parameter sets for
user-specific devices. The query returns the data as it was downloaded to the power
sensor before.
After downloading of a new user calibration data set to the power sensor, the current
S-parameter correction settings become invalid. Safe operation of the power sensor is
only possible if the
SELect
and
STATe
commands are repeated after download. See
also:
●
[SENSe<Sensor>:]CORRection:SPDevice:STATe
●
[SENSe<Sensor>:]CORRection:SPDevice:SELect
Parameters:
<caldata>
<block_data>
CALibration:USER:DATA:LENGth?
Queries the length of the user calibration data block.
Usage:
Query only
Calibrating, zeroing