![background image](http://html1.mh-extra.com/html/rands/1419-0006k02/1419-0006k02_user-manual_774879129.webp)
Remote control commands
R&S
®
NRPxxS(N)
129
User Manual 1177.5079.02 ─ 15
Example:
SYST:ERR:CODE?
Query
0
Response: No errors have occurred since the error queue was
last read out.
Usage:
Query only
SYSTem:ERRor:COUNt?
Queries the number of entries in the SCPI communication error queue.
Example:
SYST:ERR:COUN?
Query
1
Response: One error has occurred since the error queue was
last read out.
Usage:
Query only
SYSTem:ERRor[:NEXT]?
Queries the SCPI communication error queue for the oldest entry and removes it from
the queue.
Returns an error number and a short description of the error.
Example:
SYST:ERR?
Query
0, 'no error'
Response: No errors have occurred since the error queue was
last read out.
Usage:
Query only
SYSTem:FWUPdate
<fwudata>
Loads new operating firmware into the power sensor. Rohde
&
Schwarz provides the
update file. For further details, see
If you want to integrate a firmware update function in an application, see the example
given in
Chapter 7.2.3, "Using remote control"
Setting parameters:
<fwudata>
<block_data>
Definite length arbitrary block data containing the direct copy of
the binary
*.rsu
file in the following format:
#
Single digit indicating how many digits follow to specify the size
of the binary file.
Number that specifies the size of the binary file.
Binary data
Configuring the system