![background image](http://html1.mh-extra.com/html/rands/1419-0006k02/1419-0006k02_user-manual_774879137.webp)
Remote control commands
R&S
®
NRPxxS(N)
137
User Manual 1177.5079.02 ─ 15
Restarts the network connection to the DUT, that means terminates the connection and
sets it up again.
Example:
SYST:COMM:NETW:REST
Usage:
Event
SYSTem:COMMunicate:NETWork:RESet
Effective only for the R&S
NRP LAN power sensors.
Resets the LAN network settings to the default values.
Usage:
Event
SYSTem:COMMunicate:NETWork:STATus?
Effective only for the R&S
NRP LAN power sensors.
Queries the network configuration state.
Example:
SYST:COMM:NETW:STAT?
Query
UP
Response: The network is active.
Usage:
Query only
SYSTem:COMMunicate:NETWork[:COMMon]:DOMain
<domain>
Effective only for the R&S
NRP LAN power sensors.
Sets the domain of the network.
Parameters:
<domain>
Example:
SYST:COMM:NETW:COMM:DOM 'ABC.DE'
Sets
ABC.DE
as domain of the network.
SYSTem:COMMunicate:NETWork[:COMMon]:HOSTname
<hostname>
Effective only for the R&S
NRP LAN power sensors.
Sets the individual hostname of the power sensor.
In a LAN that uses a DNS server (domain name system server), you can access each
connected instrument using a unique hostname instead of its IP address. The DNS
server translates the hostname to the IP address. Using a hostname is especially use-
ful if a DHCP server is used, as a new IP address can be assigned each time the
instrument is restarted.
Configuring the system