- 16 -
Switching from weighing with the accuracy of the
II weighing range
to
weighing with the accuracy of the
I weighing range
takes place automatically
upon unloading the weighing pan and returning to AUTOZERO
(
pictogram/marker is displayed). II weighing range pictogram/marker gets blank.
11.6. Units
<P9.Unit>
parameter group enables change of start unit and temporary unit.
Unit [g] can be changed to some other unit during the weighing process or
during operation of other modes. 'Parts counting' and 'Percent weighing' modes
are exceptions.
11.6.1. Start Unit
Parameter for setting unit that is displayed and used after device start-up.
Procedure:
•
Enter
<P9.Unit / 9.1.UnSt>
submenu.
•
Press
key, available units are displayed successively one by one.
Options in case when the main unit is [kg]:
•
kg (kilogram),
•
lb (pound)*,
•
N
(Newton).
*)
–
unit disabled for verified weighing devices.
Options in case when the main unit is [g]:
•
g
(gram),
•
ct (carat),
•
lb
(pound)*.
*)
–
unit disabled for verified weighing devices.
•
Select start unit and press
key, next go back to the home screen, to
do it press
key.
•
Upon next start-up the scale runs with set start unit.