- 37 -
•
Enter respective value and press
key to confirm, home screen is
displayed automatically along with quantity of parts loaded onto the
weighing pan (pcs).
If the value of entered single part mass is greater than max
capacity value, then message <Err Hi> is displayed.
21.3. Setting Sample Mass by Determining Mass of a Single Part
•
Enter
<3.2.PcS / 3.2.2.UUt>
submenu, set
<
S_S
>
value.
•
Enter
<PcS>
working mode (parts counting), blinking value of sample
quantity is displayed.
•
Press
key to select one of the following options:
10
Reference sample quantity: 10 pcs.
20
Reference sample quantity: 20 pcs.
50
Reference sample quantity: 50 pcs.
100
Reference sample quantity: 100 pcs.
0000
Custom reference sample quantity - enter the required value yourself.
•
Select respective option and press
key to confirm, first, text
<LoAd>
is displayed for 1 second, then the weighing window.
•
If the parts are to be weighed in a container, first put the container on a
weighing pan and tare it.
•
Load the weighing pan with declared amount of parts. When the indication
is stable (
pictogram is displayed), press
key to confirm the
mass.
•
Single part mass is calculated automatically, next quantity of parts (pcs) is
displayed.
Total weight value of all parts loaded onto the weighing pan
cannot be greater than the max capacity value.
Single part mass value must be equal or greater than 0.1 of
the reading unit. Unless this condition is met, the weighing
device displays a message <Err Lo>.