![RADWAG WTC 200 User Manual Download Page 54](http://html1.mh-extra.com/html/radwag/wtc-200/wtc-200_user-manual_3904862054.webp)
- 54 -
29. ERROR MESSAGES
- E r r 2 -
- Value beyond zero range.
- E r r 3 -
- Value beyond tare range.
- E r r 4 -
- Adjustment weight or start mass out of range (
±
1% for adjustment weight,
±
10
for start mass).
- E r r L o -
- Determined mass of single part in 'Parts counting' mode too small.
- Value of 'Min' threshold is greater than value of 'Max' threshold in '+/- control'
mode.
- E r r H i -
- Entered value of single part greater than maximum capacity in 'Parts counting'
working mode.
- Entered value of 'Max' threshold greater than maximum capacity in '+/- control'
mode.
- Entered reference mass greater than maximum capacity in 'Percent weighing'
mode.
- E r r 8 -
- Time of the following operations exceeded: taring, zeroing, start mass
determining, adjustment process.
- n u l l -
- Zero value from converter.
- F U L L -
- Weighing range exceeded.
- L H -
- Start mass error, indication out of range (-5% – +15% of start mass).
- H i -
- Display range of total mass on scale display exceeded in 'Totalizing' mode.
30. TROUBLESHOOTING
Problem Cause
Solution
Battery discharged.
Connect the scale to the mains, charge
the battery (batteries).
The scale does
not switch on
No batteries (batteries not
installed or installed incorrectly).
Check if batteries are installed correctly
(polarization).
The scale
switches off
automatically
<7.3.t1> parameter set to value
enforcing scale shut-down after
particular time interval.
Go to 'Othr' menu, set <7.3.t1> parameter
to 'nonE' value.
During switching
on, message 'LH'
is displayed
Weighing pan loaded during
switching-on process.
Unload the weighing pan. Zero indication
is displayed.