- 10 -
6.4. Battery Charge Status
The scale of standard design is equipped with an internal battery. Battery state
is signalled by
pictogram, the pictogram is displayed in the top bar of the
display.
pictogram display mode
Meaning
No pictogram
Battery charged. Regular scale operation.
Pictogram displayed continuously
Too low battery charge (the scale is about to shut
down).
Charge the battery immediately.
Blinking pictogram, blink frequency:
ca. 1 s
Battery charge in progress. The device is connected to
the power supply charging the battery.
Blinking pictogram, blink frequency:
ca. 0.5 s
Battery error. Battery is damaged.
6.5. Battery Charge Status Check
•
Press
and
keys combination.
•
Battery charge state, given in
%
, is displayed for 2 s.
•
Next, home screen is displayed automatically.
7. MAINTENANCE ACTIVITIES
In order to ensure safety in the course of cleaning, it is necessary to disconnect
the device from the mains. With this condition met, uninstall the weighing pan
and other detachable components.
Cleaning the weighing pan while still installed may cause
damage of the measuring system.
7.1. Cleaning ABS Components
To clean dry surfaces and avoid smudging, use clean non-colouring cloths
made of cellulose or cotton. You can use a solution of water and detergent
(soap, dishwashing detergent, glass cleaner). Gently rub the cleaned surface
and let it dry. Repeat cleaning process if needed.
In the case of hard to remove contamination, e.g. residues of adhesive, rubber,
resin, polyurethane foam etc., you can use a special cleaning agents based on
a mixture of aliphatic hydrocarbons that do not dissolve plastics. Before using
the cleanser for all surfaces we recommend carrying out tests. Do not use
cleansers containing abrasive substances.