- 34 -
The table presents special function number and name for each of the working
modes. Remaining specific functions referring directly to a given working mode
are described further down this user manual.
19.2.1. Working Mode Accessibility
To enable/disable given working mode, press
key.
Procedure:
•
Enter
<P3.Func>
menu and select given working mode.
•
Go
to
<Acc>
function.
•
Press
key, parameter values are displayed successively one by one:
YES
– working mode enabled,
no
– working mode disabled.
•
Press
key to confirm, next go to the home screen.
19.2.2. Save Mode
Parameter allowing you to set mode of sending data from the weighing device
to a peripheral device.
Procedure:
•
Enter
<P3.Func>
menu and select given working mode.
•
Go
to
<Snn>
function.
•
Press
key, parameter values are displayed successively one by one:
StAb
Manual printout of stable weighing result. Upon pressing
key at the
moment when the result is unstable (no
pictogram displayed), the
program first waits for the stability condition to be met, only then printout is
carried out.
nStAb
Manual printout of each weighing result. In case of unstable indication,
<?>
sign is displayed in front of the 'mass frame'.
Function available for non-
verified scales exclusively
.
rEPL
Automatic printout of the first stable weighing result above
<Lo>
threshold
(to set
<Lo>
threshold go to
<Lo>
parameter).
rEPLi
Automatic printout with time interval set in
[min]
(to set the interval go to
<Int>
parameter).