- 50 -
CU0_I CR LF
- command understood but not accessible at this moment.
CU0_A CR LF
- command understood and carried out.
27.4.13. Lock Scale Keypad
Format:
K1 CR LF
Response options:
K1_I CR LF
- command understood but not accessible at this moment.
K1_OK CR LF
- command carried out.
Command is not kept in memory upon weighing device
restart.
27.4.14. Unlock Scale Keypad
Format:
K0 CR LF
Response:
K0_OK CR LF
- command carried out.
27.4.15. Give Scale Serial Number
Format:
NB CR LF
Response options:
NB_A_”Serial No.” CR LF
- command understood, response: serial number.
NB_I CR LF
- command understood but not accessible at this
moment.
”Serial No.”
– serial number of the device, inserted in between inverted commas.
Example:
NB CR LF
– command sent from a computer.
NB_A_”123456” CR LF
– scale serial number - 123456.
27.4.16. Send All Implemented Commands
Format:
PC CR LF
Response:
PC_- >_Z,T,OT,UT,S,SI,SU,SUI,C1,C0,CU1,CU0,K1,K0,NB,PC
- command carried out, the indicator displays all implemented commands.