Print the marked report on a connected printer by pressing
,
•
Using button
view the selected report.
53. APPENDIX 08 – Cooperation with HRP Weighing Platform
For detailed information on
HRP
weighing platform refer to
„HRP Weighing
Platform”
user manual. Cooperation of
PUE HY10
indicator with
HRP
weighing platform is carried out via RS232 or Ethernet interfaces using
appropriate cables that come standard with the equipment.
Upon establishing connection between
HRP
platform and
PUE HY10
indicator,
the following parameters are enabled in
<
Weighing>
submenu:
•
External ambient conditions.
•
Weighing instrument internal adjustment.
•
LO threshold for automatic functions.
53.1. External ambient conditions
Parameters allowing you to adjust your weighing device to workstation ambient
conditions.
Median Filter
The median filter is intended for eliminating short-lasting
mechanical shocks. Accessible settings:
None
(median filter is off),
0.5
,
1
,
1.5
,
2
,
2.5
.
Filter
Enter this parameter to adjust your weighing device to ambient
conditions. The higher filter level, the longer the indication takes to
stabilise. Available values:
Slow
,
Average
,
Fast
.