The computer software records the measuremnt. If P2.3.7
AUTOMATING TARRING
parameter is set to value <YES> than the
indication is automatically tared. If P2.3.7
AUTOMATING TARRING
parameter is set to value <NO> than the user must press TARE
button in order to tare the indication before (s)he doses another
porcion of water.
•
Summary of series of measurements is displayed upon its
completion.
•
Press ENTER to continue measuring process, press Esc. to stop it.
Upon pressing Esc. button the computer software returns to the main
window.
•
Upon pressing ENTER, while displaying the summary result for the
last adjustable pipette volume, a report is generated. The report is
printed on a printer connected to the balance (ambient conditions
visible on a report are those that user entered at the beginning of the
calibration procedure). For fixed volume pipette the situation is similar
if the user presses ENTER while displatying summary for volume V1.
•
The computer software returns to the main window.
•
Now the user may either start new procedure for the same pipette or
enter new data for a different pipette.