![RADWAG APP 10.R2 User Manual Download Page 117](http://html1.mh-extra.com/html/radwag/app-10-r2/app-10-r2_user-manual_781075117.webp)
- 117 -
F button shortcuts
The function enables the option of quick access for weighing functions,
which are available when buttons F1, F2, F3 or F4 are pressed.
In the mode <
DENSITY OF SOLIDS
> and <
DENSITY OF LIQUIDS
>
there are options which a user can assign to each of the
F
buttons:
NONE
– no function assigned to the button
START
– function initiates the density determination option
SELECT PRODUCT
– function initiates the option of selecting a product
from product database
LOG IN
– function initiates the option of selecting and logging-in of a
balance’s user
INSERT TARE
– function initiates the option of inserting packaging
mass
SELECT TARE
– function initiates the option of selecting packaging
mass from product database
PRINT HEADER
– function initiates the print-out of a designed header
PRINT FOOTER
– function initiates the print-out of a designed footer
VARIABLE 1
– function initiates the selection and editing of variable 1
VARIABLE 2
– function initiates the selection and editing of variable 2
The way of declaring the function is described in the section relating to
the weighing mode settings <F button shortcuts>.
17.7.1.
Density determination of solids
Density determination of solids can be carried out in two pre-defined types
of liquids or user-defined liquid with specified density:
•
WATER
(distilled water),
•
ETHANOL
(spirit 100% +/- 0.1% in temp. 20
0
C),
•
ANOTHER
(another liquid with specified density).
In case of determining density in water or alcohol it is necessary to specify
their temperature. For another liquid, its value (density) is inserted from
balance keyboard. Density determination is carried out by weighing a