4. UNPACKING, ASSEMBLING AND START-UP
4.1. Assembling and place of use
•
The balance should be stored and used in locations free of vibrations and
shakes, free of air movement and dust,
•
ambient air temperature should not exceed the range of: +10 °C ÷ +40 °C,
•
ambient relative humidity should not exceed 80%,
•
during balance operation, ambient temperature in the weighing room should
not change more than 3°C within one hour,
•
the balance should be located on a stable wall console desk or a stable
working table which is not affected by vibrations and distant from heat
sources,
•
take special precaution when weighing magnetic objects, as part of the
balance is a strong magnet. Should such loads be weighed, use under hook
weighing option, which removes the weighed load from area influenced by
the balance’s magnet. For assembling the hook for under hook weighing
option see the bottom section of balance’s housing,
•
in order to avoid the influence of static electricity on the measurement
process, ground the balance’s housing. The grounding bolt is located at the
back of the balance’s housing.