![RADWAG APP 10.R2 User Manual Download Page 59](http://html1.mh-extra.com/html/radwag/app-10-r2/app-10-r2_user-manual_781075059.webp)
- 59 -
13.1. Internal adjustment
Adjustment process can be initiated automatically and manually.
Manual means of activating adjustment procedure is achieved by pressing
. System of automatic adjustment will carry out the process fully
automatically and will inform the user on successive process stages.
Cycle of automatic adjustment process:
-
Balance software detects the necessity of carrying out adjustment and
signals it by displaying a thermometer pictogram and <Cal> at the top of
the display.
-
During this time interval of about 2 minutes within which weighing
procedure can be completed
-
As the 2 minute time elapses, balance display indicates message
CAL_30
and starts counting down from 30..29..28 to 0 (indicated value is the
counter),
-
Balance user has 30 seconds to make a decision
In order to start adjustment, do not take any actions
In order to complete weighing procedure, press
Esc.
When pressed,
balance returns to weighing procedure and displays last weighing
result. In about 5 minutes balance indicates
CAL_30
message again
.
-
The adjustment process can be postponed for multiple times, but it needs
to be pointed out that postponing of adjustment for a long time may lead to
larger errors of weighing process. The errors are the effect of temperature
changes and as a consequence changes of balance sensitivity.