- 39 -
Manual tare determination
Procedure:
•
While in optional mode press quick access button
F
, to which the
option <
ENTER TARE
> is assigned in <
SHORTCUTS
> settings
(
description of the setting further in this user manual, point <F button
shortcuts>),
•
which opens a window,
•
use the arrow buttons to enter tare value and press
button,
•
the balance returns to weighing mode, and the display indicates
entered tare value with minus „
–
" sign.
Selection of tare from TARE DATABASE
Procedure:
•
while in any working mode, press F – quick access button, to which
option <
SELECT
TARE
> is assigned in <
SHORTCUTS
>
(description of the setting further in this user manual, point <F button
shortcuts), or choose option <
SELECT TARE
> after pressing
button,
•
which displays the first mass value of the mass packages that was
saved in tare database,
•
use the arrow buttons to select the tare which is to be recalled and
press
button,
•
the balance returns to the weighing mode, and the display indicates
entered tare value with minus „
–
" sign.
Or
•
While in any working mode press
button,
•
Enter b3 option <
TARE
>
•
Which displays the first mass value of the mass packages that was
saved in tare database,
•
Use the arrow buttons to select the tare which is to be recalled and
press
button,
•
The balance returns to the weighing mode, and the display indicates
entered tare value with minus „
–
" sign.