- 103 -
17.4. Dosing
The dosing is the mode which consits in the process of sample weighing
until target mass is reached. The target mass is defined along with dosing
tolerance. Tolerance value is set as a percentage of the target mass,
entering the value percentage.
Example:
Target weight = 100.000g
Tolerance = 2,5% (2,5% of 100g, which amounts to 2,5g)
i.e.: the software accepts correctly dosed value within the following
thresholds: from 97,500g to 102,500g.
F button shortcuts
The function enables the option of quick access for weighing functions,
which are available when buttons F1, F2, F3 or F4 are pressed.
In the mode <
DOSING
> the following options are available, which a user
can assign to each of the
F
buttons:
NONE
– no function assigned to the button
ENTER SAMPLE
– function initiates the option of entering the reference
mass along with tolerance (one procedure)
SELECT PRODUCT
– function initiates the option of selecting a product
from product database
LOG IN
– function initiates the option of selecting and logging-in of a
balance’s user
ENTER TARE
– function initiates the option of inserting packaging mass
SELECT TARE
– function initiates the option of selecting packaging
mass from product database
PRINT HEADER
– function initiates the print-out of a designed header
PRINT FOOTER
– function initiates the print-out of a designed footer
VARIABLE 1
– function initiates the selection and editing of variable 1
VARIABLE 2
– function initiates the selection and editing of variable 2
The way of declaring the function is described in the section relating to
the weighing mode settings <F button shortcuts>.