4-9
4.1.10
Totalize A by B Test
4.1.10.1
Set the 2251A to its home state. Select Totalize A by B as the measurement mode,
50 ohms input A input impedance and external frequency standard.
(Command String: IP TA ALI B1)
If Option 11 is installed connect the 2251A internal reference output to the external
signal source 10 MHz Input. Configure the external signal source to accept the
10MHz reference from the 2251A.
(Command String: IP TA ALI)
4.1.10.2
Connect the 2251A to the signal generator as shown in Figure 4-7. Set the signal
generator to 10 MHz, 0 dBm. Verify that the counters measurement result is 10000
± 1.
Figure 4-7, Totalize A by B Test Configuration
4.1.10.3
Disconnect the test equipment.