3-12
3.8
Numerical Input Data
3.8.1
Table 3.9 provides the various numerical input ranges for each function and feature for
the 2251A.
Table 3.9- Numerical Input Ranges
Function
Command
Code
Numerical Limits
Low
High
Set Resolution
SRS
3
10
(Number of Digits)
Trigger Level (Xl)
SLA, SLB
-5.1
+5.1
Trigger Level (X10)
SLA, SLB
-51
+51
Math Constant
SMX, SMY,
$
1 x 10
<1 x 10
Delay/Gate Time
SDT/SGT
$
200
F
s
#
99.999s
Set-up Memory (RAM)
SM/RM
$
0
#
9
SMZ
$
-1 x 10
#
-1 x 10
-9
10
10
-9
NOTES:
1.
Entered numbers will be rounded up before storage as
follows:
a.
Trigger Level X1 to next multiple of 20 mV
b.
Trigger Level X10 to next multiple of 200 mV
c.
Delay/gate time to next multiple of 25.6
F
s
2.
Resolution entries are rounded down to the next integer.
Refer to Table 3.10 for resolution numbers and related gate
times.
3.
Math constant Z can be set to 0. However, an error message
will result if the Math function is enabled with this value set.
4.
Set-up Memory (RAM) commands require a single-digit
suffix ranging from 0 to 9.