4-4
If Option 11 is installed connect the 2251A internal reference output to the
external signal source 10 MHz Input. Configure the external signal source
to accept the 10MHz reference from the 2251A.
4.1.6.4
Connect the Signal Generator to the 2251A as shown in Figure 4-2.
Figure 4-2, Input A High Frequency Sensitivity Test Configuration
4.1.6.5
Set the 2251A measurement resolution to the values given in Table 4-2 and apply the
frequencies given to the counters input A. Verify that the 2251A gives a stable
reading within the tolerances given at less than the signal level given.
Frequency
Signal level
Resolution
Tolerance
200 MHz
50 mV rms*
9
± 1 Hz
160 MHz
50 mV rms
9
± 1 Hz
100 MHz
25 mV rms
9
± 1 Hz
10 MHz
25 mV rms
8
± 1 Hz
100 kHz
25 mV rms
8
± 0.1 Hz
* Valid up to 30
(
C
Table 4-2, Input A Sensitivity High Frequency Performance Limits
4.1.6.6
Disconnect the test equipment.