4-13
Figure 4-11, Input B Trigger Level Test Configuration
4.1.13.7
Reverse the polarity of the DC input to the counter, trigger the measurement and
recall the stored trigger level value. Verify that it is -5.000 V ± 80 mV.
4.1.13.8
Disconnect the test equipment.