DON Cutoff 0.50 ppm <0.5ppm
Result
◄
►
16
°C
18.01.09 B 09:01 pm
◄►
Print Next
PASS
DON Cutoff 0.50 ppm 0.70 ppm
Result
◄
►
16
°C
18.01.09 B 09:01 pm
◄►
Print Next
FAIL
Please note that the batch will be actualized only within the actual method (e.g. DON quant).
It is recommended to actualize all methods, as soon as a new lot is used.
Caution
It is essential to actualize always the batch number
before using a new batch. Otherwise a wrong
calculation of the results will occur.
When you have done this, select "Start Scan" and press the ENTER key to start the test strip
measurement.
While the reader is operating and scanning a RIDA
®
QUICK test strip the internal detector
moves across the test strip. Consequently, during the measurement the test strip must not be
moved. You can see on the display when a measurement is being performed. The data is
then saved, see also Sect. 3.4
Caution
When you insert the test make sure that it sits correctly
and horizontally in the proper position with the written
side facing upward. If it is at an angle, this may
damage the detector or the integrated drive system.
Depending on the chosen method, one of the following result reports is displayed after the
scanning process:
Test successful, test below limit value:
Test successful, test above limit value:
Set Cut Off measured result
Result report
Figure 12 a
Figure 12 b