Chapter 2: Quick Tutorial
6
.
This quick tutorial provides step-by-step instructions that will take
you through the basic operating procedures and give you a general
understanding of how the IAQ Monitor is used.
If you have used the IAQ Monitor before, or if you are comfortable
using instrumentation, this will probably give you the information
you need to get started. Refer to the remainder of this reference
manual for detailed instructions.
This tutorial is very straightforward and will be easy to follow even if
you have never used the IAQ Monitor and are not familiar with
technical instrumentation. It will help get you acquainted with the
operating procedures you will be using.
NOTE:
The CO
2
, temperature and relative humidity sensors are
built into the IAQ Monitor. See "Connecting and Calibrating Your
Sensors" in Chapter 4 for information on connecting the toxic gas
and linear sensors.
Turning the IAQ Monitor On and Off
1. Press the ON/OFF button. The IAQ Monitor will turn on and
automatically begin displaying the present readings.
2. Press the ON/OFF button again. The following screen will
appear:
Press RECORD
To Turn Off
ON/OFF to Continue
3. Press RECORD. The unit will turn off.
NOTE:
In order to turn off the IAQ Monitor it CANNOT be
recording.
Calibrating the CO
2
Sensor
1. Unscrew and remove the toxic sensor cap (see the IAQ Monitor
instrument diagram in the beginning of Chapter 3).
2. Slide the calibration adapter over the wand until the "O" ring in
the bottom of the adapter slides over the handle of the wand.