SERVICE MANUAL
Rev. 05/2005
P - 55
The adjustments of the chapters 8-2-1 and 8-2-2 are
interactive.
8-2-2 AIMING BEAM FOCALIZED AND SHARP
CAUTION :
Do not change the focus distance of the slit lamp (locked).
NOTE :
When you move the laser head, you move the Yag beam and aiming
beam in the same time.
Search the position of the laser head where the aiming beam is the
more sharp as possible and where the two beams are superim-
posed, to do one aiming beam.
Standard Gap between Slit Generator
and Objective of biomicroscope : 3mm
ENERGY
ENERGY
VISEE
FOCUS
FOCUS
This screw is used to adjust the focalization of the
aiming beam. You can improve the focus.
16
25
10
16
2
0
1
Standard Gap
Correct position of
the slit lamp
Bad position of
the slit lamp
Before touching the above screws,
you must untighten two screws on
one side of the laser head.
NOTE :
The final adjustment must be obtained when all screws are
tightened.
8 - OPTICAL ADJUSTMENTS