SERVICE MANUAL
Rev. 05/2005
P - 34
4 - INSTALLATION
S T OP
a) Place the chinrest and screw it on the table with the two long six-point
socket screws.
b) Insert a metal shaft in the hole at the side of cross-slide table.
Position two pieces of gears at both ends of the metal shaft and fix it with
six-point socket screws, placing a groove between each gear and groove.
c) Place the slit lamp on the table.
Position the gears correctly on the rack of the table to avoid malfunction of
the cross-slide table.
Make certain that bottom sides (viewed from operator) of the
cross-slide table are parallel to the side-lines of white plate. In the
opposite case the table can not move.
Good position
d) Place the slit generator axis on the slit
lamp axis and screw it.
Gear
Gear
Metal shaft
Cross slide table
Groove
Screw
Groove
Screw
Bad position
+
S T OP
CAUTI O N : FEDERAL USA L AW RESTRI CTS THI S DEVI CE TO SAL E BY O R O N THE
O RDER O F A PHYSI CI AN O R AN O PTO M ETRI ST L I CENCED BY THE L AW O F THE
STATE I N W HI CH HE PRACTI SES TO USE O R O RDER THE USE O F THI S DEVI CE
DANG ER : EXPLO SI O N HAZARD. DO NO T USE I N PRESENCE O F
FLAM M ABLE ANESTETI CS O R O XI G EN. RI CH ATM O SPHERE
CAUTI O N : ELECTRI C SHO CK HAZARD. DO NO T O PEN. REFER
SERVI CI NG TO Q UALI FI ED SERVI CE PERSO NNEL.
DANGER
LASER RADI ATI O N
Avoi d eye or ski n exposure t o
di rect or scat t ered radi at i on
Nd : YAG 1064 nm 30mJ M ax
( 3 p u l s e s 4
Di ode 637 nm 1 mW
CLASS I V LASER PRO DUCT