SERVICE MANUAL
Rev. 05/2005
P - 35
4 - INSTALLATION
I
0
Use only with a 250 V fuse
Employer uniquement avec
un fusible de 250 V
+
+
QUANTEL MEDICAL
21, rue Newton, Z.I. du Brézet,
63039 Clermont-Ferrand cédex 2
FRANCE
Tel : +33 (0) 473 745 745 - Fax : + 33 (0) 473 745 7 00
REF :
SN :
OPTIMIS II
THIS PRODUCT
COMPLIE S WITH 21 CFR 10 40
CAUTION :
U.S. F EDERAL LAW
RESTRICTS THIS
OR ON ORDER
OF A PHYSICIAN
!
IP 20
4 AT
250 V
50/60 Hz
150 VA, 1
Φ
100 - 230V~
AP
G
MADE IN FRANCE
0459
LASER C LASS 4/IV
N d:YAG Las er : 1064nm,
30mJ Max , (3 puls es x 4ns )
VISIBLE AND/OR INVISIBLE LASER RADIATION
AVOID E YE OR SK IN EXPOS URE TO
DIRECT OR SCATTERED RADIATION
LASER C LASS 2/II
D iode Las er : 637nm, <1mW Max
CL ASS IV L ASER PRODUCT
I
0
Use only with a 250 V fuse
Employer uniquement avec
un fusible de 250 V
+
+
Slit
Lamp
Joystick
Footswitch
Interlock
H.V.
Voltage
Viridis
Interlock
GND
GROUND
SN :
51 02 02
STOP
f) Connect the following cables and remote :
- Sub-D 25 cable
- Power supply cable
- Slit Lamp cable
- Joystick interlink cable or footswitch cable
- High voltage cable
- Remote
- Ground cable
g) Close the rear panel with the large screws passing the cables in the holes
on the rear panel.
High voltage
connector
Main
switch
Ground
connector
Hole threaded
to screw knurled
Sub-D 25 pins
connector
Joystick or footswitch
connector
Remote
connector
Slit lamp power
supply connector
Large screws to keep in place rear panel
e) Place the both gear covers sliding it under the rail guides of the cross slide
table.
Gear cover
Gear cover
Rail guide
Rail guide
Hole to pass the cables
Hole to pass the cables
Hole threaded
to screw knurled