QIAsymphony SP/AS Consolidated Operating Guide 12/2017
161
Error code
Description
Comments and suggestions
45008
Selected assay(s) and a number of
assay points cannot be used as there
are no adapters available that can
hold the required amount of reagent
and/or number of assay control
tubes.
Ensure that all available adapters are listed in the
Adapters AS
list. Configure the correct number of available adapters in the
Configuration
menu, see Section 6.1.3 of the
QIAsymphony
SP/AS User Manual — General Description
for more
information.
To reduce the number of tube positions required on the reagent
holders, reduce the number of assays or reduce the number of
samples. Alternatively you could reduce the number of replicates
for samples, assay controls, assay standards, and no template
controls.
45101
There is not enough space on all
assay racks to handle the defined
samples, assay controls, and assay
standards.
Reduce the number of selected samples, assay standards,
assay controls, and no template controls.
You could also reduce the number of replicates for samples,
assay controls, assay standards, and no template controls.
45103
There are not enough adapters (X)
available for labware type Y. Check
the adapter configuration of your
QIAsymphony AS.
Ensure that all available adapters are listed in the
Adapters AS
list. Configure the correct number of available adapters in the
Configuration
menu, see Section 6.1.3 of the
QIAsymphony
SP/AS User Manual — General Description
for more details
about how to do this.
If there are no more adapters of that type available, reduce
the number of replicates for samples, assay standards, and
assay controls.
If the error persists, contact QIAGEN Technical Services.
70110
A disk space overrun was
recognized. The current usage is X.
Do you want to delete result files
older than Y days?
The system detected that there is less than 15% free disk
space. Result files older than 10 days (default) can be deleted
from the instrument(s). Press
Yes
to free up file space and
delete older result files.
Note
: If you choose to delete results files, they cannot be
recovered. Be sure to download result files older than 2 days
to the USB stick before you decide to free up disk space.
70118
Bar code reading error on slot X. Try
insertion again and reduce insertion
speed!
A bar code reading error occurred. A sample tube may be
incorrectly positioned, a bar code may be dirty, or a bar code
may not be clearly printed.
Either, correct the position of the sample tube, clean the bar
codes, check the bar code printout, or reduce the speed at
which the sample tubes are loaded.
70401
Some of the combinations of ACS
and sample labware that you
selected are marked as “not
recommended” [List of not
recommended combinations].
Press
No
to stay in this dialog and
assign different ACS or go to the
previous dialog to change the sample
labware. Do you want to continue
anyway?
The selected sample tubes/rack are not recommended for use
with the Assay Control Set.
Perform one of the following:
Press
Yes
to ignore the message.
Press
No
and change the sample tube/rack.
Use the
Labware Browser
menu or the labware lists to identify
which labware is recommended for use with which protocols.
Labware lists are available for download at
www.qiagen.com/goto/QIAsymphony.
The “Supervisor” can go to the
Configuration
menu and
disable the parameter
Check combination of protocol and
recommended labware during run definition?
in the
Process
SP 2
tab.
70402
Sample X: Labware Y is not
recommended for use with ACS Z.
The sample tube/rack is not recommended for use with the
Assay Control Set.
See error code 70401.
70504
It was not possible to define all
batches automatically. You must
define them separately.
Not all sample batches could be defined using the
Fast Setup
button. Define each batch separately. If sample errors occurred,
these will be displayed during definition of the relevant batch.
Summary of Contents for QIAsymphony AS
Page 213: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 213 This page intentionally left blank...
Page 214: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 214 This page intentionally left blank...
Page 216: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 216 www qiagen com...