QIAsymphony SP/AS Consolidated Operating Guide 12/2017
187
14.4.6
Robotic gripper (SP)
1.
Wipe the robotic gripper with a lint-free cloth moistened with ethanol-based disinfectant (e.g.,
Mikrozid). Incubate as appropriate.
2.
Wipe with a lint-free cloth moistened with water and dry with paper towels.
Important
Only wipe the weight. Do not wipe the rods otherwise the ball mechanism may
become jammed.
14.4.7
Liquid waste container (SP)
1.
Remove the liquid waste container from the “Waste” drawer.
2.
Empty the liquid waste container. Dispose of the liquid waste according to your local safety
regulations.
3.
Clean the liquid waste container with a glyoxal and quaternary ammonium salt based
disinfectant (e.g., Gigasept Instru AF) according to the manufacturer’s instructions.
4.
Rinse the liquid waste container with deionized water.
5.
Replace the liquid waste container in the “Waste” drawer.
14.5
Weekly maintenance (SP/AS)
14.5.1
File management
1.
Download the result file(s) (for QIAsymphony SP and QIAsymphony AS) and loading
information files (QIAsymphony AS only) as described in Section 6.3 and ensure that the files
are backed up.
2.
Delete result files older than 10 days (default setting) as described in Section 6.6.
14.5.2
Cleaning the touchscreen
Wipe the touchscreen with ethanol-based disinfectant (e.g., Mikrozid). Then wipe with a cloth
moistened with water and dry with paper towels.
Summary of Contents for QIAsymphony AS
Page 213: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 213 This page intentionally left blank...
Page 214: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 214 This page intentionally left blank...
Page 216: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 216 www qiagen com...