To program the unit for
Example 2 Level Monitoring and Control
by using
the
Quick Setup
menu proceed as follows. If required access the
Program
Mode
, key in the
passcode
1997
and press
ENTER.
At the
Quick Setup
menu press the hot key and toggle to Point 2 display
and press
ENTER
and then as prompted, by the questions, select the
relevant option and
ENTER
.
Using the ‘right’ arrow key, go to the
Quick Setup
menu press
ENTER
and
as prompted, by the questions, select the relevant option and press
ENTER
.
Programming is now complete, and the unit can be returned to the run
mode, press
CANCEL
until
Run Mode?
Is displayed on the LCD press
ENTER
, and the UltraTWIN will return to the
Run Mode
.
QUESTION
OPTION
Level/Volume
1 = Level App.
Control
1 = Control down
No. of controls
1 = 1 Relay
Control no.1
1 = Set to relay 1
No. of alarms
2 = 2 Alarms
Type alarm 1
1 = High
Alarm no.1
1 = Set to relay 2
Type alarm 2
2 = Low
Alarm no.2
5 = Set Relay 5
Xducer (P*101)
2 = dB6
Material (P*102)
1 = Liquid
Measurement units (P*104) 4 = feet
Empty Level (P*105)
11.0 feet
Span (P*106)
10.0 feet
Important Notice
If relay setpoints do
n’
t meet the exact requirements of the application, they
can be modi
fied to suit by pressing ENTER when, “For More Options Hit
Enter”, is displayed and entering new values to relay setpoints as required.
Alternatively, the relevant relay setpoint can be accessed either by the main
menu system or directly via parameter number and changed, as necessary.