Page
54
Response
125
0 = Slow
1 = Fast
Slow
Changing this to ‘Fast’ will
automatically calibrate
parameters in the MicroFlow
sensor to track measurements
faster. This is recommended
for Pumped flow. ‘Slow’ is
recommended when there is
natural flow, as
measurements will be tracked
at a slower pace.
Multipoint
Calibration @
0.2 m/s
152
5 - 500
100
In percentage terms, default
is 100, for 100%. 50%
would half and 200% would
double.
Multipoint
Calibration @
0.4 m/s
153
5 - 500
100
In percentage terms, default
is 100, for 100%. 50%
would half and 200% would
double.
Multipoint
Calibration @
0.8 m/s
154
5 - 500
100
In percentage terms, default
is 100, for 100%. 50%
would half and 200% would
double.
Multipoint
Calibration @
1.5 m/s
155
5 - 500
100
In percentage terms, default
is 100, for 100%. 50%
would half and 200% would
double.
Multipoint
Calibration @
3.0 m/s
156
5 - 500
100
In percentage terms, default
is 100, for 100%. 50%
would half and 200% would
double.
Multipoint
Calibration @
5.0 m/s
157
5 - 500
100
In percentage terms, default
is 100, for 100%. 50%
would half and 200% would
double.
Multipoint
calibration
mode
158
0 = Off
1 = On
Off
Turns Multipoint calibration
on and off.
Summary of Contents for MICROFLOW
Page 1: ...MICROFLOW INSTRUCTION MANUAL...
Page 2: ......
Page 4: ......
Page 6: ...58 59...
Page 12: ...Page 6 EU Declaration of Conformity...
Page 31: ...Page 25 Distance Distance Diagram 4 Diagram 3 Distance Distance Distance...
Page 36: ...Page 30 This page is left blank intentionally...
Page 44: ...Page 38 This page is left blank intentionally...
Page 56: ...Page 50 This page is left blank intentionally...