Page 49
The option for manually setting the parameters can be done by utilising the Manual
Setting terminal on the parameters tab, as shown below:
All parameters can be queried by entering the parameter address in the Parameter
field and then clicking on the Query button. The
MicroFlow
will then reply with
the current value of that parameter.
For setting a parameter, both the Address and Value must be entered before
clicking on the Set button. ‘User’ will show the value you have changed in the
manual setting box. And ‘UI’ will appear when changing a parameter via one of
the parameter drop down selection boxes. The appropriate access levels are
required for setting parameters. After changing any parameter, the ‘SAVE’ button
must be clicked to send and save the new values to the device. If the change is not
successful, then the values that are shown on the Parameter tab will reflect the
current parameter value on the sensor.
The values need to be valid and within the allowed range for that particular
parameter for the change to be successful. Please refer to ‘
Chapter 7
Parameter
listing and Descriptions
’ in this manual, on the parameters available and the
range each one can be set between. If a parameter has been incorrectly set,
highlighting the ‘Default’ box and selecting any of the options from the drop down
list will display that particular parameters default box. This shows the parameter
number, its current value and the default value of that individual parameter. Select
‘OK’ and then ‘SAVE’ to store that value.
If the MicroFlow is being controlled simultaneously on both
MicroFlow PC
and
RS485 via Modbus interface, the values on the Parameter tab will not be instantly
updated on MicroFlowPC if a parameter change is instructed via the RS485
Modbus.
Summary of Contents for MICROFLOW
Page 1: ...MICROFLOW INSTRUCTION MANUAL...
Page 2: ......
Page 4: ......
Page 6: ...58 59...
Page 12: ...Page 6 EU Declaration of Conformity...
Page 31: ...Page 25 Distance Distance Diagram 4 Diagram 3 Distance Distance Distance...
Page 36: ...Page 30 This page is left blank intentionally...
Page 44: ...Page 38 This page is left blank intentionally...
Page 56: ...Page 50 This page is left blank intentionally...